05 LHP 151versusLTQ Orbitrap XL Mass

Melles Griot 05 LHP 151, Thermo Fisher LTQ Orbitrap XL Mass, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
05 LHP 151Melles Griot
OEMMelles GriotThermo Fisher
CategoryPrecision OpticsPrecision Optics
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
FamilyMelles Griot 05Thermo Fisher LTQ Orbitrap XL Mass
Specifications
ConfigurationWith 05 LPL 340 Power Supply (Does not power on)[1]IonMax ESI Source[3]
Equipment typeLaser[1]—
MakeMelles Griot[1]Thermo Scientific[3]
Model05 LHP 151[1]LTQ Orbitrap XL[3]
Power stateDoes not power on[1]—
Max Resolution—100,000[3]
Ion source—IonMax ESI Source[3]
Mass spectrometer dimensions—141.4 × 87 × 146.3 cm (H × W × D)[4]
Mass spectrometer weight—~600 kg[4]
Power requirements—230 Vac ±10% 3 phase, 16 Amps, 50/60 Hz, with earth ground for the instrument[4]
Gas supply—One high purity (99.5% pure, flow rate 15 L/min) nitrogen gas supply for the API source; one ultra-high-purity helium gas supply (99.998%) with less than 1 ppm each of water, oxygen, and total hydrocarbons for the linear ion trap[4]
Maximum resolution—100,000[3]
Resolution (apodized)—60,000 (FWHM) at m/z 400 with a scan repetition rate of 1 second[5]
Minimum resolution—7,500 at m/z 400[5]
Mass range—m/z 50–2000; m/z 200–4000[5]
Mass accuracy—<3 ppm RMS for 2 h period with external calibration using defined conditions; <2 ppm RMS with internal calibration[5]
Dynamic range—>10000 between mass spectra; >4000 between highest and lowest detectable mass in one spectrum[5]
Collision cell—HCD collision cell[5]
Recirculating chiller—Thermo Scientific ThermoFlex 900[5]
Instrument type—Hybrid mass spectrometer[5]
Main components—Linear ion trap, intermediate storage device (curved linear trap), Orbitrap analyzer, collision cell[5]
Resolution range—7,500 to 100,000 at m/z 400[5]
Scan functions—MS/MS and MSn[5]
HCD collision cell—Included[4]
Power—230 Vac ±10% 3 phase, 16 Amps, 50/60 Hz, with earth ground for the instrument; 120 or 230 Vac single phase with earth ground for the data system; 120 or 230 Vac single phase, 15 Amps, with earth ground for the water chiller[4]
Weight—~600 kg[4]
Dimensions—141.4 × 87 × 146.3 cm (H × W × D)[4]

Plan your fab

Building a Precision Optics process? Get a complete equipment list.

Open the fab equipment planner →
Sources (5)Every fact above is drawn from these public sources
  1. [1]inventory listing
  2. [2]waferpedia.comwaferpedia.com
  3. [3]inventory listing
  4. [4]documents.thermofisher.comdocuments.thermofisher.com
  5. [5]docs.thermofisher.comdocs.thermofisher.com