The Perkin Elmer AAnalyst 400 is an atomic absorption spectrometer that features a true double-beam Echelle optical system with a solid-state detector. The AAnalyst 400 covers a spectral range from 190 to 900 nm. The AAnalyst 400 supports analysis techniques including flame, graphite furnace, and hydride generation systems.[1][2][3]

AAnalyst 400 Atomic Absorption — Inventory photo
Fig. 01AAnalyst 400 Atomic AbsorptionInventory photo[4]

Power

100-230 V (±10%), 50/60 Hz, 300 VA maximum[3]

Optics

190 to 900 nm[1]

What it is

The Perkin Elmer AAnalyst 400 Atomic Absorption is an atomic absorption spectrometer for elemental analysis.[4][1][2]

The AAnalyst 400 is a laboratory analytical instrument in the atomic absorption class.[5][3]

How it works

The AAnalyst 400 uses atomic absorption optics and software control to measure elements by light absorption.[3][2]

The AAnalyst 400 uses a true double-beam Echelle optical system, a segmented solid-state detector, and automatic lamp alignment.[3][2]

The AAnalyst 400 supports flame, graphite furnace, and hydride or flow-injection methods through modular hardware and software control.[1][3][6]

Where it fits in the process flow

The AAnalyst 400 is used for elemental measurement in routine analysis and in more demanding applications.[1][4][5]

Applications

The AAnalyst 400 is used for elemental analysis across flame, furnace, and hydride techniques.[1][3]

The AAnalyst 400 is suitable for measuring trace elements and corrosive or high-solid sample matrices.[1][2][3]

What do the numbers mean?

Power & electrical1

Power Requirements
100-230 V (±10%), 50/60 Hz, 300 VA maximum[3]
Accurate?

Optics & imaging6

Spectral Range
190 to 900 nm[1]
Accurate?
Detection Range
Covers the spectral range from 190 to 900 nm[1]
Accurate?
Optical System
True double-beam echelle optical system[3]
Accurate?
Detector
High-efficiency, segmented solid-state detector[3]
Accurate?
Wavelength Range
189-900 nm[3]
Accurate?
Spectral Bandpass
0.15 nm at 200 nm[3]
Accurate?

Dimensions & facilities3

Sample Compartment Dimensions
25 cm wide x 25 cm deep[3]
Accurate?
Dimensions (W x D x H)
70 x 65 x 65 cm[3]
Accurate?
Weight
49 kg[3]
Accurate?

Configuration & options10

Sensitivity
Achieves detection limits in the parts per billion (ppb) range[1]
Accurate?
Focal Length
300 mm[3]
Accurate?
Grating
36 x 185 mm area, 79 lines/mm, blaze angle 76°[3]
Accurate?
Prism
Fused-quartz prism: 95 x 40 mm, 60°[3]
Accurate?
Light Sources
Hollow cathode or electrodeless discharge lamps (EDLs); four-lamp turret[3]
Accurate?
Background Corrector
Deuterium background corrector[3]
Accurate?
Operating Temperature Range
+15 °C to +35 °C[3]
Accurate?
Humidity Range
20% to 80% non-condensing[3]
Accurate?
Burner Head (Air/Acetylene)
10-cm single-slot solid titanium burner head included[3]
Accurate?
Spray Chamber
Inert-polymer mixing chamber[3]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Power Requirements100-230 V (±10%), 50/60 Hz, 300 VA maximum[3]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What optical system does the AAnalyst 400 use?

The AAnalyst 400 uses a true double-beam Echelle optical system.[3][2]

What lamp types does the AAnalyst 400 accept?

The AAnalyst 400 accepts hollow cathode lamps and electrodeless discharge lamps.[3][2]

What analysis techniques can the AAnalyst 400 support?

The AAnalyst 400 supports flame, graphite furnace, and FIAS techniques.[3][1]

What software is used with the AAnalyst 400?

The AAnalyst 400 uses WinLab32 for AA software.[1][2][3]

Not publicly documented

The following facts about the AAnalyst 400 Atomic Absorption are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the AAnalyst 400 Atomic Absorption are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the AAnalyst 400 Atomic Absorption are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the AAnalyst 400 Atomic Absorption are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the AAnalyst 400 Atomic Absorption are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the AAnalyst 400 Atomic Absorption is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (6)Every fact above is drawn from these public sources
  1. [1]americanlaboratorytrading.com — americanlaboratorytrading.comamericanlaboratorytrading.com
  2. [2]http://www.bilmar.com.tr/PDF/PERKIN%20ELMER/B-PE-AAS400.pdf — bilmar.com.trbilmar.com.tr
  3. [3]richmondscientific.com — richmondscientific.comrichmondscientific.com
  4. [4]Inventory photo
  5. [5]AAnalyst 400 Atomic Absorption Spectrophotometer, Perkin Elmer - STEMart — ste-mart.comste-mart.com
  6. [6]Perkin Elmer AAnalyst 400 Atomic Absorption Spectrometer with HGA 900 | Express Lab Werks, LLC — expresslabwerks.comexpresslabwerks.com
Was this page accurate?

Last updated Oct 2, 2026.