Waferpedia

Keysight

E 5250 A Low Leakage Switch Mainframe

Provisional entry — research in progress

This page was generated automatically from cited public sources and hasn't completed the full research pass yet. Every specification shown carries its source; more detail is added as research completes.

E 5250 A Low Leakage Switch Mainframe — Inventory photo
Fig. 01E 5250 A Low Leakage Switch MainframeInventory photo[1]

What it is

A low leakage switch mainframe is a programmable switching system used in semiconductor parametric test to route signals between measurement instruments and devices under test. The Keysight E 5250 A is a low leakage switch mainframe that expands a single measurement station to an automated measurement system.[2][3]

How it works

The E 5250 A switches DC current, DC voltage, capacitance, and pulse signals with DC instruments and LCR meters. The mainframe controls switching through a built-in GPIB interface. Plug-in modules can be configured either as a cross-point matrix for general parametric measurements or as a multiplexer for long-term reliability measurements.[2][3]

Switch modules plug into four slots that accept 30 mm height switch cards. The mainframe performs a self-test using the optional relay function test adapter (Option 301). Typical performance includes an offset current of less than 0.1 pA on the low leakage I-V port and a relay contact life greater than 10^8 operations in dry switching mode.[2][3]

What do the numbers mean?

Power & electrical5

Power requirements
100/120/220/240 V ± 10%, 47 Hz to 63 Hz[2]
Accurate?
Max voltage rating (E5252A) - channel to ground
200 V[2]
Accurate?
Max voltage rating (E5252A) - channel to channel
300 V[2]
Accurate?
Typical offset voltage (EMF) at 5 min (E5252A) - Low Leakage I-V Port
< 80 μV[2]
Accurate?
Bandwidth (-3 dB) (C-V, HF Port)
< 10 MHz[2]
Accurate?

Dimensions & facilities4

Number of slots
4 slots for 30-mm height switch cards[2]
Accurate?
Weight (mainframe)
11.2 kg[2]
Accurate?
Weight (E5252A)
2.0 kg[2]
Accurate?
Weight (E5255A)
2.1 kg[2]
Accurate?

Configuration & options25

Maximum VA
100 VA[2]
Accurate?
Size (mainframe)
230 mm H x 430 mm W x 600 mm D[2]
Accurate?
Size (E5252A card)
30 mm H x 395 mm W x 424 mm D[2]
Accurate?
Size (E5255A card)
30 mm H x 395 mm W x 424 mm D[2]
Accurate?
Operating temperature range
5°C to 40°C[2]
Accurate?
Storage temperature range
-40°C to 70°C[2]
Accurate?
Operating humidity range
5% to 80% (no condensation)[2]
Accurate?
Storage humidity range
5% to 90% RH at 65°C[2]
Accurate?
Altitude (operating)
0 to 2000 m (6500 ft)[2]
Accurate?
Altitude (storage)
0 to 15,240 m (50,000 ft)[2]
Accurate?
No. of ports (with E5252A matrix)
Low leakage I-V Port: 2, General I-V Port: 4, C-V Port: 2, HF Port: 2, Output Channels: 12[2]
Accurate?
No. of ports (with E5255A multiplexer)
Low leakage I-V Port: 3, Bias Port: 3, Output Channels: 24 (8 x 3)[2]
Accurate?
Max current rating (E5252A)
1.0 A[2]
Accurate?
Close channel residual R (E5252A) - Low Leakage I-V Port
0.6 Ω[2]
Accurate?
Close channel residual R (E5252A) - General I-V Port
1.0 Ω[2]
Accurate?
Close channel residual R (E5252A) - C-V, HF Port
1.0 Ω[2]
Accurate?
Channel isolation R (E5252A) - Low Leakage I-V Port (Condition A)
1e13 Ω[2]
Accurate?
Channel isolation R (E5252A) - General I-V Port (Condition A)
1e12 Ω[2]
Accurate?
Channel isolation R (E5252A) - C-V, HF Port (Condition A)
1e9 Ω[2]
Accurate?
Typical offset current (E5252A) - Low Leakage I-V Port
< 0.1 pA[2]
Accurate?
Typical offset current (E5252A) - General I-V Port
< 1000 pA[2]
Accurate?
Typical channel crosstalk capacitance (E5252A)
< 0.3 pF/ch[2]
Accurate?
Typical guard capacitance (E5252A) - Low Leakage I-V Port
< 145 pF[2]
Accurate?
Relay contact life (dry switching mode)
> 10^8 times[2]
Accurate?
Settling time (Low Leakage I-V Port)
< 3.5 sec / 0.4 pA[2]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Power requirements100/120/220/240 V ± 10%, 47 Hz to 63 Hz[2]
  • Max voltage rating (E5252A) - channel to ground200 V[2]
  • Max voltage rating (E5252A) - channel to channel300 V[2]
  • Typical offset voltage (EMF) at 5 min (E5252A) - Low Leakage I-V Port< 80 μV[2]
  • Bandwidth (-3 dB) (C-V, HF Port)< 10 MHz[2]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What plug-in modules are available for the E 5250 A?

The E 5250 A accepts the E5252A 10 x 12 matrix switch and the E5255A 24-channel (8 x 3) multiplexer. Up to four modules can be installed per frame.[2][3]

What are the power requirements?

The E 5250 A requires 100/120/220/240 V ±10%, 47 Hz to 63 Hz, with a maximum VA of 100 VA.[2][3]

What is the typical offset current on the low leakage I-V port?

The typical offset current is less than 0.1 pA when zero volts are applied to all input and output channels.[2][3]

What is the relay contact life?

The relay contact life is greater than 10^8 operations in dry switching mode.[2][3]

What is the operating temperature range?

The operating temperature range is 5°C to 40°C with a humidity range of 5% to 80% (no condensation).[2][3]

Not publicly documented

The following facts about the E 5250 A Low Leakage Switch Mainframe are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the E 5250 A Low Leakage Switch Mainframe are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the E 5250 A Low Leakage Switch Mainframe are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the E 5250 A Low Leakage Switch Mainframe are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the E 5250 A Low Leakage Switch Mainframe are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the E 5250 A Low Leakage Switch Mainframe is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (3)Every fact above is drawn from these public sources
  1. [1]Inventory photo
  2. [2]testunlimited.com — testunlimited.comtestunlimited.com
  3. [3]https://www.transcat.com/media/pdf/keysight-E5250A-datasheet.pdf — transcat.comtranscat.com
Was this page accurate?