| Attribute | Dionex UltiMate 3000Thermo Fisher | Dionex UltiMate 3000 Liquid ChromatographyThermo Fisher |
|---|---|---|
| OEM | Thermo Fisher | Thermo Fisher |
| Category | — | — |
| Wafer size | — | — |
| Process node | — | — |
| Introduced | — | — |
| Production run | — | — |
| Lifecycle | — | — |
| Lifecycle milestones | — | — |
| Control system | — | — |
| Generation | — | — |
| Family | Thermo Fisher Dionex UltiMate | Thermo Fisher Dionex UltiMate |
| Cited variants | ||
| Specifications | ||
| Configuration | Autosampler[3] | High performance[6] |
| Manufacturer | Thermo Scientific / Dionex[3] | — |
| Model | UltiMate 3000[3] | UltiMate 3000[6] |
| System components listed | Autosampler, RS Pump, Column Compartment, Tray[3] | — |
| Pump family | UHPLC, HPLC, and biocompatible pumps[4] | — |
| Pump flow range | 20 nL/min to 10 mL/min[4] | — |
| Maximum pressure | Up to 1250 bar[4] | — |
| RSLCnano flow range | 20 nL/min up to 50 μL/min[5] | — |
| RSLCnano maximum pressure | 800 bar[5] | — |
| RSLCnano gradient delay volume | 25 nL[5] | — |
| RSLCnano sample injection | High-precision sample injections down to 10 nL[5] | — |
| Solvent reservoir capacity | — | Eight 1-L, four 2.5-L, two 5-L, or combinations thereof, up to sixteen 0.5-L reservoirs[7] |
| Wetted parts | — | PEEK, FEP, and Tefzel[7] |
| Inline filter frit | — | Stainless steel (SST) or Titanium[7] |
| Environmental conditions | — | Indoor use; temperature 10-35 °C; humidity ≤80% non-condensing; pollution degree 2[7] |
| Dimensions (solvent rack) | — | 10 x 42 x 51 cm (h x w x d)[7] |
| Weight (solvent rack) | — | 3.0 kg without reservoirs[7] |
| Pump types | — | SD, RS, BM[2] |
| Typical purge flow rate (analytical pumps) | — | 3 mL/min[2] |
| Typical purge flow rate (micro pumps) | — | 2 mL/min[2] |
| Upper pressure limit (RS pumps) | — | 75 MPa[2] |
| Upper pressure limit (SD pumps) | — | 45 MPa[2] |
| Upper pressure limit (BM pumps) | — | 35 MPa[2] |
| Make | — | Thermo Scientific / Dionex[6] |
| Description | — | Liquid Chromatography[6] |
| Application | — | HPLC analysis[7] |